  <dc:title>A high sensitivity method for the determination of nanomolar concentrations of nitrate and nitrite in seawater with a technicon autoanalyzer 2</dc:title>
  <dc:creator>/Oudot, Claude</dc:creator>
  <dc:creator>/Montel, Yves</dc:creator>
  <dc:subject>NITRATE</dc:subject>
  <dc:subject>NITRITE</dc:subject>
  <dc:subject>EAU DE MER</dc:subject>
  <dc:subject>ANALYSE AUTOMATIQUE</dc:subject>
  <dc:description>Un manifold haute sensibilit&#xE9; pour la d&#xE9;termination des traces (concentrations nanomolaires) de nitrate + nitrite et nitrite seul est d&#xE9;crit. La m&#xE9;thode utilise un AutoAnalyseur TECHNICON 2 classique, employ&#xE9; habituellement pour les analyses &#xE0; bord des bateaux. Une reproductibilit&#xE9; de + ou - 1 nmol dm-3 pour N03 + N02 seul &#xE9;tait obtenu avec une cadence d'analyse de 4O &#xE9;chantillons par heure. (R&#xE9;sum&#xE9; d'auteur)</dc:description>
  <dc:date>1988</dc:date>
  <dc:type>text</dc:type>
  <dc:identifier>https://www.documentation.ird.fr/hor/fdi:26640</dc:identifier>
  <dc:identifier>fdi:26640</dc:identifier>
  <dc:identifier>Oudot Claude, Montel Yves. A high sensitivity method for the determination of nanomolar concentrations of nitrate and nitrite in seawater with a technicon autoanalyzer 2. 1988, 24,  239-252</dc:identifier>
  <dc:language>EN</dc:language>
</oai_dc:dc>